Device product

Alinity i Processing Module, (a) Part Number 03R65-01 (b) Part Number A-30108552-01 (c) Part Number A-30109694-01 Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.

Z-0789-2019

December 20, 2018

Class II

Product summary

Firm
Abbott Gmbh & Co. KG
Event
Event 81916
Status
Terminated
Classification
Class II
Quantity
434 units
Official record key
device-enforcement:Z-0789-2019

Official wording

Reason: Abbott has identified an issue with the Alinity i Gear Pump Assembly (part number A 30108552 01) resulting in foaming/bubbling out of the bottle reservoir for Concentrated Wash Buffer and an unexpected amount of dried residue of buffer. The foaming/bubbling is a result of an improper calibration of the pump s operating speed.

Code information: (a) Serial Numbers: Ai01001 - Ai01455, Ai01457, Ai01459 (b) all Serial Numbers (c) all Serial Numbers

Distribution pattern: Worldwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Abbott has identified an issue with the Alinity i Gear Pump Assembly (part number A 30108552 01) resulting in foaming/bubbling out of the bottle reservoir for Concentrated Wash Buffer and an unexpected amount of dried residue of buffer. The foaming/bubbling is a result of an improper calibration of the pump s operating speed.