Device product

NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi .0.8.521 9+PO1 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Z-0773-2019

January 09, 2019

Class II

Product summary

Firm
NEUSOFT MEDICAL SYSTEMS IMP & EX
Event
Event 81920
Status
Terminated
Classification
Class II
Quantity
3
Official record key
device-enforcement:Z-0773-2019

Official wording

Reason: For surview scan length more than 500mm,if the user aborts or skips the reconstructed surview, the image may have the wrong field of view (FOV). When user plans and performs subsequent clinical scan with the wrong FOV the actual start and end position may not be consistent with the planned positons.

Code information: Series number includes: N1281180242E, N1281170231E, N1281180248E

Distribution pattern: US Distribution in states of: KS, OK and MA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    For surview scan length more than 500mm,if the user aborts or skips the reconstructed surview, the image may have the wrong field of view (FOV). When user plans and performs subsequent clinical scan with the wrong FOV the actual start and end position may not be consistent with the planned positons.