Skip to content
Recall Observatory FDA recall evidence

Device product

ULRICH Torque Limiting Handle for use with a Set Screw Driver to perform the final tightening of set screws in a spinal implant construct.

Z-1126-2019

March 08, 2019

Class II

Product summary

Firm
ulrich medical USA Inc
Event
Event 82497
Status
Terminated
Classification
Class II
Quantity
25
Official record key
device-enforcement:Z-1126-2019

Official wording

Reason: Certain Torque Limiting Handles in the field are beyond their 3 year calibration life.

Code information: Model #s CS 7130 Serial #s 538436-02_70 (69 total) UDI Description 4052536123704, Model #s CS 7134 Serial #s 538438-02_70 (69 total) UDI Description 4052536123728, Model #s CS 7146 Serial #s 04-14-022-18_23 (6 total) UDI Description4052536078301, Model #s CS 7148 Serial #s 11-12-035-01_09, 03-12-006-24, and 03-12-006-25 (11 total) UDI Description 4052536078318, and Model #s CS 8049 Serial #s J04192-31_40, J04192-056, J04192-064, H12015-031_6, and J01276-031_3, (21 total) UDI Description 4052536022076.

Distribution pattern: AZ, IL, LA, MI, MO, NV, OH, PA, TN, and TX.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Certain Torque Limiting Handles in the field are beyond their 3 year calibration life.