Skip to content
Recall Observatory FDA recall evidence

Device product

Randox CALCIUM (Ca) Colorimetric Method RX Series Cat. No. CA 3871, R1. Arsenazo Reagent 9 x 51ml GTIN: 05055273200904

Z-0839-2019

December 04, 2018

Class II

Product summary

Firm
Randox Laboratories Ltd.
Event
Event 81746
Status
Terminated
Classification
Class II
Quantity
225 kits
Official record key
device-enforcement:Z-0839-2019

Official wording

Reason: Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.

Code information: All lot codes

Distribution pattern: WV, PR

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.