Skip to content
Recall Observatory FDA recall evidence

Device product

aap Implante AG, Lambotte chisel, 20 mm, REF IU 3000-20 Product Usage: Surgical instruments are intended to aid in the insertion of bone plates and screws as well as for reduction and support during orthopedic surgeries. If not otherwise specified, instruments are intended for multiple use.

Z-1019-2019

January 11, 2019

Class II

Product summary

Firm
AAP Implantate Ag
Event
Event 82159
Status
Terminated
Classification
Class II
Quantity
25 units
Official record key
device-enforcement:Z-1019-2019

Official wording

Reason: There is the potential for the Lambotte chisel to break at the welded piece parts of the chisel body and knocking plate.

Code information: Lot Numbers: I011, I012, I013, I014, I015, I016, I017

Distribution pattern: Worldwide Distribution - US Nationwide in the state of Texas and country of Germany

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0