Device product
CERTAIN(R) BELLATEK(R) TIN ABUTMENT 3.4MM, Item Number ILDAN3 - Product Usage: BIOMET3i Restorative products are intended for use as an accessory to endosseous dental implants for placement in the maxilla and mandible
Z-2143-2019
Product summary
- Firm
- Biomet 3i, LLC
- Event
- Event 82263
- Status
- Terminated
- Classification
- Class II
- Quantity
- 2 units
- Official record key
device-enforcement:Z-2143-2019
Official wording
Reason: During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.
Code information: Lot Numbers/UDI: 8442561-2/(01)00844868031352(10)8442561-2 8442561-3/(01)00844868031352(10)8442561-3
Distribution pattern: Nationwide and Puerto Rico, Canada, and Australia
Derived failure modes
-
Unknown
During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.