Device product
AURIGA 30 GENERAL SYSTEM UPN: M0068S30G0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.
Z-1397-2019
Product summary
- Event
- Event 82691
- Status
- Terminated
- Classification
- Class II
- Quantity
- 18 (OUS)
- Official record key
device-enforcement:Z-1397-2019
Official wording
Reason: Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.
Code information: Serial Number: 21302793, 21302835, 21302799, 21302844, 21302861, 21302780, 21302921, 21302999, 21303007, 21303013, 21303012, 21303064, 21303032, 21303033, 21303045, 21303065, 21302816, 21302994
Distribution pattern: Worldwide distribution to Australia, Belgium, Brazil, Chile, Colombia, Egypt, France, Germany, Great Britain, India, Indonesia, Iran, Italy, Lebanon, Malaysia, Mexico, Romania, Russia Fed., South Africa, South Korea, Spain, Taiwan, United Arab Emirates and Vietnam
Derived failure modes
-
Unknown
Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.