Vigilant Agilia, Vigilant Drug'Lib, REF Z073476, versions 1.0 and 1.1 Product Usage: Vigilant Drug Lib is Dose Error Reduction Software that, when used in combination with Volumat MC Agilia infusion pumps, is intended to reduce drug administration errors at the bedside. Vigilant Drug Lib Agilia is a software application running on a computer and intended to configure and generate multiple device configurations, drug libraries, care areas and data sets that can then be uploaded to the Volumat MC Agilia infusion pumps. Intended patient population: The Vigilant Drug Lib software must only be used to create drug libraries for patients whose weight is 250g to 250 kg (0.6 lbs. to 551 lbs.).

Z-2065-2019

June 24, 2019

Class I

Product summary

Firm
Fenwal Inc
Event
Event 83138
Status
Terminated
Classification
Class I
Quantity
83 units
Official record key
device-enforcement:Z-2065-2019

Official wording

Reason: The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC Agilia Software and versions 1.0 and 1.1 of Vigilant DrugLib

Code information: All serial numbers distributed in the US

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software anomalies