Device product
Siemens Artis Q floor --Interventional Fluoroscopic X-Ray System Model Number: 10848280
Z-1898-2019
Product summary
- Event
- Event 83082
- Status
- Terminated
- Classification
- Class II
- Quantity
- 12
- Official record key
device-enforcement:Z-1898-2019
Official wording
Reason: Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel
Code information: Serial Numbers: 103008 103030 103028 103258 103221 103222 103277 103220 103279 103285 103286 103295
Distribution pattern: Nationwide
Derived failure modes
-
Unknown
Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel