Device product

LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components - LEEP PRECISION Cart, LEEP PRECISION Generator, and the LEEP PRECISION Smoke Evacuator for loop electrosurgical excision procedures. The loop electrosurgical excision procedure (LEEP) utilizes a very thin wire in the shape of a loop heated by electrical current generated by a power supply to efficiently remove abnormal cells. The LEEP PRECISION INTEGRATED SYSTEM has been designed to provide an efficient workstation that can efficiently meet all the unique LEEP requirements of the clinic and practice-based facilities .

Z-1840-2019

April 23, 2019

Class II

Product summary

Firm
CooperSurgical, Inc.
Event
Event 82821
Status
Terminated
Classification
Class II
Quantity
19
Official record key
device-enforcement:Z-1840-2019

Official wording

Reason: 2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses.

Code information: 2018110001,2018110002, 2018110003, 2018110004, 2018110005,2018110006, 2018110007, 2018110008,2018110011, 2018110012,2018110013, 2018110014, 2018110015,2018110016, 2018110017, 2018110019, 2018110020,2018110021,& 2018110022

Distribution pattern: US Nationwide Distribution in the states of CA, CT, IN, MA, MI, MN, NC, NV, NY, OH, TX, WA and WI.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0