Device product

AquaBeam Handpiece, Sterile EO, Rx Only, REF: 310301, and 320301, a component of AquaBeam Robotic System. 320301 distributed outside the U.S. Device Description The AQUABEAM Robotic System is intended for use in patients suffering from lower urinary tract symptoms resulting from benign prostatic hyperplasia (BPH).

Z-1110-2019

March 06, 2019

Class II

Product summary

Firm
PROCEPT BIOROBOTICS CORPORATION
Event
Event 82462
Status
Terminated
Classification
Class II
Quantity
214
Official record key
device-enforcement:Z-1110-2019

Official wording

Reason: Certain lots may develop an internal leak due to a micro-crack in a specific weld joint within the device. When the Robot detects this failure, it immediately renders the device inoperable and notifies the user via an error message, which may cause delay and/or inability to complete the procedure.

Code information: Domestic Lots: 18C00681, 18C00770, 18C00852, 18C01244. Foreign Lots: 18C00799, 18C00844, 18C00985, 18C01024, 18C01044, 18C01234

Distribution pattern: Worldwide Distribution: US (nationwide) state to: TX, FL, IN, CA, IL, VA, MI, PA.; and (OUS) countries of: Italy, Germany, France, Poland, United Kingdom, Spain, Canada and Australia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Certain lots may develop an internal leak due to a micro-crack in a specific weld joint within the device. When the Robot detects this failure, it immediately renders the device inoperable and notifies the user via an error message, which may cause delay and/or inability to complete the procedure.