Device product

Monaco RTP System, software versions 5.10 and 5.11, Radiation Treatment Planning System

Z-2091-2019

July 03, 2019

Class II

Product summary

Firm
Elekta Inc
Event
Event 83273
Status
Ongoing
Classification
Class II
Quantity
1396 units
Official record key
device-enforcement:Z-2091-2019

Official wording

Reason: When creating 3D plans using either MU or Dose weighting modes, if the user changes the number of fractions, rescales the plan and then changes the wedge angle, the Monitor Units are scaled incorrectly.

Code information: all Monaco Radiation Treatment Planning units installed with software version 5.10 or 5.11

Distribution pattern: worldwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    When creating 3D plans using either MU or Dose weighting modes, if the user changes the number of fractions, rescales the plan and then changes the wedge angle, the Monitor Units are scaled incorrectly.