Philips Azurion systems with software version R1.2 -Interventional Fluoroscopic X-ray system as follows: Azurion 3 M12, Azurion 3 M15, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20 Models: 722063, 722064, 722067, 722068, 722078, 722079. Product Usage: The Azurion series (within the limits of the used Operating Room table) are intended for use to perform: Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures. Additionally: . The Azurion series can be used in a hybrid Operating Room. The Azurion series contain a number of features to support a flexible and patient centric procedural workflow.

Z-2205-2019

July 16, 2019

Class II

Product summary

Firm
Philips North America, LLC
Event
Event 83477
Status
Terminated
Classification
Class II
Quantity
834
Official record key
device-enforcement:Z-2205-2019

Official wording

Reason: The cold restart of Azurion R1.2 systems may take up to 7 minutes if the system is connected to the mains power supply for more than 50 days, may result in a delay of treatment.

Code information: Software version R1.2

Distribution pattern: US Nationwide Distribution Foreign: Canada Argentina AUSTRALIA Austria Azerbaijan Bahrain Bangladesh Belgium Bosnia and Herzegovina Brazil Bulgaria Chile Colombia CROATIA Denmark ECUADOR Egypt Finland France Germany Greece Hong Kong India Indonesia Ireland Israel Italie Italy Japan Jordan Kazakhstan Korea, Republic of Latvia Lebanon Libya Malaysia Maldives Mexico Mongolia Myanmar Netherlands New Zealand Norway Oman Pakistan Philippines Polamnd Poland Portugal Republic of Korea Romania Russian Federation Saudi Arabia Serbia Singapore SLOVENIA South Africa Spain Sri Lanka Sweden Switzerland Thailand Tunisia Turkey UNITED ARAB EMI United Arab Emirates United Kingdom Uzbekistan Viet Nam

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The cold restart of Azurion R1.2 systems may take up to 7 minutes if the system is connected to the mains power supply for more than 50 days, may result in a delay of treatment.