Device product
Microdot Xtra blood glucose test strips
Z-2357-2019
Product summary
- Event
- Event 82867
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 17,784 (USA)
- Official record key
device-enforcement:Z-2357-2019
Official wording
Reason: The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.
Code information: LOT NUMBERS 7091101, 7091201, 7091801, 7091902, 7100402, 7100403, 7101002, 7101602, 7101702, 7102403, 7103002, 7110602, 7110703, and 7112702.
Distribution pattern: Illinois
Derived failure modes
-
Unknown
The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.