Device product

Artis zee biplane MN, Model Number 10094143

Z-2033-2019

June 12, 2019

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 83203
Status
Terminated
Classification
Class II
Quantity
3
Official record key
device-enforcement:Z-2033-2019

Official wording

Reason: A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Code information: Serial Numbers: 149202 149100 149106

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.