Device product

Alinity c Cuvette Segments, List Number 04S47-01 Product Usage: The Alinity c Cuvettes are used on the Alinity c system processing module. The Alinity c processing module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and potentiometric detection technology.

Z-1709-2019

April 15, 2019

Class II

Product summary

Firm
Abbott Gmbh & Co. KG
Event
Event 82644
Status
Terminated
Classification
Class II
Quantity
19,184
Official record key
device-enforcement:Z-1709-2019

Official wording

Reason: Individual cuvettes within the Alinity c Cuvette Segment may become seated lower than the designed height. This may result in inadequate dispense into specific cuvettes due to the sample probe being unable to make efficient contact with the cuvette bottom.

Code information: All serial numbers

Distribution pattern: US Nationwide Distribution- AL, AR,CA, FL, GA, IL, LA, MA, ME, MN, MO, NJ, NY, OK, SC, TN, TX, UT, & VA WORLDWIDE - Albania, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Columbia, Croatia, Czech Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, Norway, Pakistan, Philippines, Poland, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Thailand, Trinidad & Tobago, Turkey, United Kingdom, and Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Individual cuvettes within the Alinity c Cuvette Segment may become seated lower than the designed height. This may result in inadequate dispense into specific cuvettes due to the sample probe being unable to make efficient contact with the cuvette bottom.