Zyno Medical Administration Set -105"ADMIN SET 20DP W/FILTER 2YSITES CLMP- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code: B2-70072-F

Z-2059-2019

May 29, 2019

Class II

Product summary

Firm
Zyno Medical LLC
Event
Event 83026
Status
Terminated
Classification
Class II
Quantity
360 cases (50/case)=18,000
Official record key
device-enforcement:Z-2059-2019

Official wording

Reason: Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified

Code information: Lot Numbers: 18046309, 18026344 UDl/UPC/GTIN: 00814371020204

Distribution pattern: MA

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified