Device product
ARIES Flu A/B & RSV Assay , REF 50-10020, UDI # 00840487100158
Z-2284-2019
Product summary
- Event
- Event 83470
- Status
- Ongoing
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-2284-2019
Official wording
Reason: Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.
Code information: Lot numbers and discontinue use by dates for Part Number 50-10020: AA7127 - Immediately; AA7206 - Immediately; AA7312 - 3/5/2019; AA7549 - 3/30/2019; AA7635 - 4/10/2019
Distribution pattern: US: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom
Derived failure modes
-
Unknown
Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.