Recall Observatory FDA recall evidence

Device product

Cardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes, with trocar Product Number: JP-2189 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.

Z-2262-2019

June 28, 2019

Class II

Product summary

Firm
Degania Silicone, Ltd.
Event
Event 83287
Status
Terminated
Classification
Class II
Quantity
6080
Official record key
device-enforcement:Z-2262-2019

Official wording

Reason: 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body

Code information: Lot Codes: P1841752 P1852268 P1845823

Distribution pattern: Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body