Device product

Elekta Unity, Image-Guided Radiation Therapy System Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.

Z-1980-2019

June 24, 2019

Class II

Product summary

Firm
Elekta, Inc.
Event
Event 83158
Status
Terminated
Classification
Class II
Quantity
12 units
Official record key
device-enforcement:Z-1980-2019

Official wording

Reason: The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.

Code information: UDI/GTIN: 5060081071321 Installed product: UNITY SYSTEM/13553-006/600003 UNITY SYSTEM/10817-011/600010 UNITY SYSTEM/12473-020/600020 UNITY SYSTEM/12274-017/600012 UNITY SYSTEM/10562-011/600016 UNITY SYSTEM/10719-T03001/600013 UNITY SYSTEM/30004070-001/600023 UNITY SYSTEM/10420-AVL-U/600008 UNITY SYSTEM/11014-UMCU-U/600007 UNITY SYSTEM/10157-006/600014 UNITY SYSTEM/11611-40/600011 UNITY SYSTEM/11489-21/600009

Distribution pattern: US Distribution to states of: TX & WI, and Internationally to: Canada, Denmark, Germany, Italy, Netherlands, Sweden, and the UK.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.