Device product

Alaris Pump, Model 8100, with bezel mechanical assemblies manufactured with FR-110 plastic resin. The following products are affected by this recall as they are pumps or pump assemblies with bezels manufactured between April 2011 and June 2017 with the FR-110 plastic. Alaris Pump Module serviced with LVP Mechanism Sub Assembly (P/N 10942012, P/N 49000007, and P/N 49000203), and Alaris Pump module Bezel Kit Assembly (P/N 10964559 and P/N 49000204). Product Usage: The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using the IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural. There are 6 bezel posts on the Alaris Pump bezel assembly.

Z-1768-2019

April 13, 2018

Class I

Product summary

Firm
CareFusion 303, Inc.
Event
Event 82143
Status
Terminated
Classification
Class I
Quantity
607618
Official record key
device-enforcement:Z-1768-2019

Official wording

Reason: The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as other damage to the bezel (i.e. external cracking). The separation of one or more bezel posts may result in free flow, over infusion, under infusion or interruption of infusion.

Code information: All pumps manufactured between April 2011 and June 2017, that have bezel mechanical assemblies manufactured with FR-110 plastic resin. And all pump assemblies with bezels manufactured between April 2011 and June 2017 with the FR-110 plastic.

Distribution pattern: Worldwide Distribution - U.S. Nationwide CA, TX, NC, AL, ID, TN, IN, MA, UT, HI, NH, PA, IL, OR, FL, MI, VA, AR, OH, GA, NY, MD, MN, NM, AZ, KY, WI, NE, MS, DC, MO, KS, SC, LA, MT, WA, NJ, CO, CT, OK, AK, AS, CU, DE, GU, IA, ME, MP, ND, NV, PR, SD, VT, WV, WY; OUS: AE, AU, BH, CA, CH, CN, GI, IL, IN, KW, MX, MY, NL, NZ, PH, PK, QA, SA, SG, TW, and ZA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as other damage to the bezel (i.e. external cracking). The separation of one or more bezel posts may result in free flow, over infusion, under infusion or interruption of infusion.