Skip to content
Recall Observatory FDA recall evidence

Device product

Olympus WM-NP2 or WM-DP2 Mobile Workstations with Nurse Control Panel Arm, Model. No. MAJ-1664

Z-1983-2019

June 10, 2019

Class II

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 83171
Status
Terminated
Classification
Class II
Quantity
N/A
Official record key
device-enforcement:Z-1983-2019

Official wording

Reason: WM-NP/DP2 workstations could become unstable and may fall over if the control panel arm is mounted with a video monitor weighing more than 19.8 pounds.

Code information: All serial numbers

Distribution pattern: Nationwide domestic distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    WM-NP/DP2 workstations could become unstable and may fall over if the control panel arm is mounted with a video monitor weighing more than 19.8 pounds.