Device product

TransWarmer Warming Infant Transport Mattress, Model no. 20421 Product Usage: The TransWarmer¿ Warming Infant Transport Mattress is gel-filled, disposable mattress that provides heat for up to two hours when cold stress is a concern. The patented WarmGel heats up to 104 (when starting from 75 ) in 60 seconds or less. The product is sold in boxes of 6 Infant Transport Mattresses per box. The intended use of the product is to provide warmth during infant transport within the hospital or between hospitals.

Z-2364-2019

July 08, 2019

Class II

Product summary

Firm
CooperSurgical, Inc.
Event
Event 83110
Status
Terminated
Classification
Class II
Quantity
49813
Official record key
device-enforcement:Z-2364-2019

Official wording

Reason: The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is prohibited and could lead to serious health consequences, such as skin burns. some products do not contain the updated IFU.

Code information: IJ780, IJ782, IJ785, IJ789, IJ794, IJ796, IJ804, IJ807, IJ817, IJ824, IJ832, IJ834, IJ841, IJ853, IJ865, IJ874, IJ878, IJ884, IJ886, IJ892, IJ895, IJ897, IJ900, IJ902, IJ909, IJ917, IJ920, IJ924, IJ933, IJ946, IJ948, IJ957, IJ962, IJ973, IJ976, IJ982, IJ985, IJ999, IK007, IK009, IK015, IK028, IK038, IK041, IK053, IK060, IK066, IK095, IK102, IK107, IK127, IK135, IK147, IK162, IK164, IK171, IK178, IK184, IK201, IK214, IK238, IK252, IK258, IK265, IK270, IK276, IK298, IK318, IK322, IK330, IK332, IK339, IK353, IK361, IK378, IK384, & IK390

Distribution pattern: Worldwide Distribution - US Nationwide in all 50 states. International distribution to Australia, Brazil, Canada, Ecuador, Germany, Gibraltar, India, Ireland, Kuwait, Macedonia, New Zealand, Oman, Poland, Puerto Rico, Qatar, Romania, Saudi Arabia, South Korea, Switzerland, United Arab Emirates, & United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is prohibited and could lead to serious health consequences, such as skin burns. some products do not contain the updated IFU.