Skip to content
Recall Observatory FDA recall evidence

Device product

Vista Dental Products Chlor-XTRA - 3mL Prefilled Syringes, QTY: 12, REF 503825

Z-0016-2020

June 27, 2019

Class II

Product summary

Firm
Inter-Med Llc
Event
Event 83435
Status
Terminated
Classification
Class II
Quantity
58 units
Official record key
device-enforcement:Z-0016-2020

Official wording

Reason: The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

Code information: UDI (01)10818207020496 Lot/Work Order Numbers: 20190565 20191107

Distribution pattern: US, South Africa, Switzerland, Poland, Italy, Singapore, Australia

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.