Device product
HS Hi-R NEO 900 Reference # 657 820 Serial # from 101 Product Usage: A wide range of surgical interventions are made possible through the modular construction and the diversity of the equipment and accessories. The operating microscope HS Hi-R NEO 900/HS Hi-R NEO 900A/HS Hi-R NEO 900A NIR and HS ALLEGRA 900/590/90 are designed for use in Ophthalmology and in other disciplines where a vertical line of viewing is required. Due to their high maneuverability, the HS ALLEGRA range of operating microscopes are qualified for further applications, in particular ENT- and reconstructive microsurgery are possible, because their line of sight can be tilted or adjusted horizontally. The HS ALLEGRA 590 is particular because of its multidisciplinary application possibilities.
Z-2482-2019
Product summary
- Event
- Event 83571
- Status
- Terminated
- Classification
- Class II
- Quantity
- 0
- Official record key
device-enforcement:Z-2482-2019
Official wording
Reason: software error in the central control unit of the floor stand.
Code information: The Error only occurs in combination with the following floor stands FS 2-21, FS 2-25, FS 3-45 with the software REF 615 586 in the versions 10.1.xx to 11.3.xx 1. Product: FS 2-21, Reference # 615 510 Serial # 101 to 360, SW-Version: 10.1.xx to 11.3.xx 2. Product: FS 2-21, Reference # 615 510, serial # from 361 3. Product: FS 2-25, Reference # 615 550, Serial # 101 to 376, SW Version: 10.1.xx to 11.3.xx 4. Product: FS 2-25, Reference # 615 550, Serial # from 377 4. Product: FS 3-45, Reference # 615 705, Serial #from 101
Distribution pattern: US in the state of OH
Derived failure modes
-
Device software or design
software error