Skip to content
Recall Observatory FDA recall evidence

Device product

Siemens Artis Q zeego ---Interventional Fluoroscopic X-Ray System Model Number: 10848283

Z-1901-2019

June 06, 2019

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 83082
Status
Terminated
Classification
Class II
Quantity
2
Official record key
device-enforcement:Z-1901-2019

Official wording

Reason: Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel

Code information: Serial Numbers: 117309 117125

Distribution pattern: Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel