Device product
Siemens Artis Q biplane---Interventional Fluoroscopic X-Ray System Model Number: 10848282
Z-1900-2019
Product summary
- Event
- Event 83082
- Status
- Terminated
- Classification
- Class II
- Quantity
- 73
- Official record key
device-enforcement:Z-1900-2019
Official wording
Reason: Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel
Code information: Serial Numbers: 121013 121057 121056 121058 121006 121185 121016 121027 121020 121321 121448 121040 121049 121015 121094 121082 121089 121383 121144 121319 121137 121085 121155 121193 121136 121165 121140 121141 121151 121168 121169 121154 121211 121188 121415 121183 121209 121187 121309 121200 121313 121206 121361 121366 121359 121393 121301 121407 121406 121352 121379 121364 121342 121471 121520 121411 121412 121402 121430 121502 121421 121472 121460 121457 121515 121444 121516 121510 121491 121525 121530 121527
Distribution pattern: Nationwide
Derived failure modes
-
Unknown
Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel