Skip to content
Recall Observatory FDA recall evidence

Device product

Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.

Z-0360-2020

February 08, 2019

Class II

Product summary

Firm
Canon Medical System, USA, INC.
Event
Event 84135
Status
Terminated
Classification
Class II
Quantity
9 systems
Official record key
device-enforcement:Z-0360-2020

Official wording

Reason: Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.

Code information: Model CAT-870B D053991 Serial No. Manufacture Date 99D08X2101 October 2008 99D0582039 August 2005 99D0762072 June 2007 99D0852090 May 2008 99D07Y2078 November 2007 99D0812081 January 2008 99D0582042 August 2005 99D0552033 May 2005 D551206 May 2005

Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of AL, CT, FL, IL, OR, PA, and TX; and country of Netherlands.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.