Skip to content
Recall Observatory FDA recall evidence

Device product

Patient Data Manager, Model/Catalogue Numbers: 25100-09 - PATIENT DATA MANAGER 2.1.1LICENSE ONLY 25100-10 - PATIENT DATA MANAGER 2.2.0 LICENSE ONLY 25100-11 - PATIENT DATA MANAGER 2.2.1 LICENSE ONLY 25100-12 - PATIENT DATA MANAGER 2.3 INSTALLER 25100-12A - PATIENT DATA MANAGER 2.3.1 INSTALLER 25100-13 - PATIENT DATA MANAGER 2.3 LICENSE 25100-12B - PATIENT DATA MANAGER 2.4.0 INSTALLER 25100-13A - PATIENT DATA MANAGER 2.4 LICENSE 25100-12C - PATIENT DATA MANAGER 2.5.0 INSTALLER 25100-12D - PATIENT DATA MANAGER 2.5.1 INSTALLER 25100-13B - PATIENT DATA MANAGER 2.5 LICENSE used in combination with iPlan CMF/Cranial/ENT/Spine 3.0.5, 3.0.6. Product Usage: Patient Data Manager is intended to load patient data from and export patient data to connected network locations (like file shares and PACS) and removable media. Furthermore, the software is intended to start other applications (like Brainlab planning and navigation applications) and to route applications and connected video sources to displays attached to the system that are running Patient Data Manager. The software is also intended to connect two Brainlab systems to each other via network.

Z-0176-2020

September 25, 2019

Class II

Product summary

Firm
Brainlab AG
Event
Event 83978
Status
Terminated
Classification
Class II
Quantity
1230 units
Official record key
device-enforcement:Z-0176-2020

Official wording

Reason: The Brainlab Patient Data Manager software (PDM) in combination with the Brainlab iPlan CMF/Cranial/ENT/Spine software 2.1.1, 2.2.0, 2.2.1, 2.3.0, 2.3.1, 2.4.0, 2.5.0, 2.5.1 (if used in combination with iPlan CMF/Cranial/ENT/Spine 3.0.5, 3.0.6) that may lead to shifts applied to segmentation objects in iPlan being not reflected when exported into DICOM under specific circumstances.

Code information: Patient Data Manager versions 2.1.1 no GTIN 2.2.0 no GTIN 2.2.1 no GTIN 2.3.0, GTIN 04056481006662 2.3.1, GTIN 04056481132453 2.4.0, GTIN 04056481138615 2.5.0, GTIN 04056481140502 2.5.1, GTIN 04056481140502 versions 2.1.1, 2.2.0, 2.2.1, 2.3.0, 2.3.1, 2.4.0, 2.5.0, 2.5.1.

Distribution pattern: worldwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The Brainlab Patient Data Manager software (PDM) in combination with the Brainlab iPlan CMF/Cranial/ENT/Spine software 2.1.1, 2.2.0, 2.2.1, 2.3.0, 2.3.1, 2.4.0, 2.5.0, 2.5.1 (if used in combination with iPlan CMF/Cranial/ENT/Spine 3.0.5, 3.0.6) that may lead to shifts applied to segmentation objects in iPlan being not reflected when exported into DICOM under specific circumstances.