Device product
Viva - JR, catalog # 6002-940, UDI # 03661540600210 Product Usage: automated chemistry analyzer intended for clinical use
Z-0154-2020
Product summary
- Event
- Event 83582
- Status
- Terminated
- Classification
- Class II
- Quantity
- 4 units
- Official record key
device-enforcement:Z-0154-2020
Official wording
Reason: The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.
Code information: serial numbers: 11-1302, 11-303, 11-304 , 11-1305
Distribution pattern: US Nationwide in the states of NY, and UT
Derived failure modes
-
Unknown
The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.