Device product
BriteBlade Pro Single-Use Fiber Optic Mac 3, CE01120, Do Not Re-use, Rx Only, REF 040-713U, Sterile EO, Qty 10, (01)15055788723001 The device is intended for direct vision laryngoscopy to obtain an illuminated view of the vocal cords and the glottis by insertion into the oral cavity to manipulate the tongue and/or the epiglottis.
Z-0801-2020
Product summary
- Event
- Event 84367
- Status
- Terminated
- Classification
- Class II
- Quantity
- 1,024,890 devices
- Official record key
device-enforcement:Z-0801-2020
Official wording
Reason: The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.
Code information: Part Number/Lot Number 040-713U 180700030 I CCW 040-713U 180900099 I CCW 040-713U 180901353 I CCW 040-713U 181000110 I CCW 040-713U 181000113 I CCW 040-713U 181100001 I CCW 040-713U 181100002 I CCW 040-713U 181200005 I CCW
Distribution pattern: US: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK
Derived failure modes
-
Unknown
The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.