Device product
OMNIlife science, Inc. ApeX acetabular liner hip prosthesis component Serf Dual Mobility Cup Liner 53-28; Product code: H6-25328
Z-2479-2019
Product summary
- Event
- Event 83528
- Status
- Terminated
- Classification
- Class II
- Quantity
- 14 units
- Official record key
device-enforcement:Z-2479-2019
Official wording
Reason: Acetabular liners failed the oxidation testing performed prior to release. They may experience a higher wear rate, leading to revision surgery.
Code information: LOT # 31706
Distribution pattern: US nationwide distribution to states of: IL, NY, OK, MD, AL, PA, CO, FL, and UT.
Derived failure modes
-
Unknown
Acetabular liners failed the oxidation testing performed prior to release. They may experience a higher wear rate, leading to revision surgery.