Skip to content
Recall Observatory FDA recall evidence

Device product

MassHunter Quantitative Analysis Software, Version B.07.01, used with Triple Quadrupole Mass Spectrometers, models K6460 and K6420 Product Usage: A mass spectrometer for clinical use is a device intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass.

Z-2359-2019

August 15, 2018

Class II

Product summary

Firm
Agilent Technologies, Inc.
Event
Event 82794
Status
Terminated
Classification
Class II
Quantity
235
Official record key
device-enforcement:Z-2359-2019

Official wording

Reason: When utilizing specific software versions of an LC/MS device, and Batch at a Glance in Compound Table view, a defect occurs when a new sample(s) is inserted. This defect creates a mismatch between the sample name and the column header whereby the sample header will be offset by one sample. The recalling firm requests that you discontinue using this workflow to generate reports because the Quant batch table will not display fully analyzed results or save any changes made.

Code information: Software versions B.07 or below

Distribution pattern: US Nationwide. distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    When utilizing specific software versions of an LC/MS device, and Batch at a Glance in Compound Table view, a defect occurs when a new sample(s) is inserted. This defect creates a mismatch between the sample name and the column header whereby the sample header will be offset by one sample. The recalling firm requests that you discontinue using this workflow to generate reports because the Quant batch table will not display fully analyzed results or save any changes made.