Device product
TxRx 15CH Knee Coil 1.5T QED part number:Q7000016 Siemens Model # 10185453
Z-0032-2020
Product summary
- Event
- Event 83775
- Status
- Terminated
- Classification
- Class II
- Quantity
- 3 units
- Official record key
device-enforcement:Z-0032-2020
Official wording
Reason: Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.
Code information: Serial # 1471 2106 1991
Distribution pattern: Worldwide distribution - US in the states: NC OUS: Germany
Derived failure modes
-
Unknown
Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.