Skip to content
Recall Observatory FDA recall evidence

Device product

Randox Liquid Cardiac Controls Catalogue Numbers CQ5053.

Z-0414-2020

October 04, 2019

Class II

Product summary

Firm
Randox Laboratories, Limited
Event
Event 84044
Status
Terminated
Classification
Class II
Quantity
6,702
Official record key
device-enforcement:Z-0414-2020

Official wording

Reason: Removal of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5053 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots.

Code information: Catalogue Number: CQ5053 GTIN: 05055273207460 Batch/lot: 4245CK (Exp. 28Nov2019), 4248CK (Exp. 28Nov2019), 4313CK (Exp. 28Jun2020) and 4316CK (Exp. 28Jun2020)

Distribution pattern: US Distribution: WV, CA, DE, NJ, NY, GA, IN, PR OUS Distribution: CA, PL, CH, JP, NL, IT, CN, GB, IE, RU, MD, BY, SA, VN, AE, BH, DK, CO, ES, FR, TW, PH, TH, OM, HU, IN, KR, ZA, AU, CZ, PE, CL, IR, NZ, BE, DE, BA, SV, SK, HK, PT, PK, BR, TR, UA, SE, QA, LT, PE,

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Removal of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5053 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots.