Skip to content
Recall Observatory FDA recall evidence

Device product

Vista Dental Products Sodium Hypochlorite - 6%, 10 - 12cc Pre-filled Syringes, REF 502300

Z-0012-2020

June 27, 2019

Class II

Product summary

Firm
Inter-Med Llc
Event
Event 83435
Status
Terminated
Classification
Class II
Quantity
1992 units
Official record key
device-enforcement:Z-0012-2020

Official wording

Reason: The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

Code information: Lot/Work Order Numbers: 2018-1348 2018-1585 2018-1586 2018-1852 20182106, UDI (01)10818207020465 20182302, UDI (01)10818207020465 20182458, UDI (01)10818207020465 20190725, UDI (01)10818207020465 20190835, UDI (01)10818207020465

Distribution pattern: US, South Africa, Switzerland, Poland, Italy, Singapore, Australia

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.