Philips DigiTrak XT Holter Recorder Software version number B.1 and C.2. The DigiTrak XT Recorder is available in four different configurations. Service tag numbers for these configurations are as follows: - 453564067151 24-Hour Configuration - 453564067161 48-Hour Configuration - 453564067171 96-Hour Configuration - 453564067181 7-Day Configuration Model Number: 860322

Z-1345-2020

January 13, 2020

Class II

Product summary

Firm
Philips North America, LLC
Event
Event 84879
Status
Terminated
Classification
Class II
Quantity
44,442 units
Official record key
device-enforcement:Z-1345-2020

Official wording

Reason: if an AAA battery is inserted in the recorder and a user attempts to start it, or if the recorder is inserted in the docking station, the recorder will display Error: 602 and fail to function for a new patient study.

Code information: All Philips 860322 DigiTrak XT Holter Recorder, manufactured or distributed before December 31, 2019, are affected, regardless of serial number, ship date, and date of manufacture.

Distribution pattern: US Nationwide Foreign: Afghanistan Albania Algeria Argentina Armenia Aruba Australia Austria Azerbaijan Bangladesh Belarus Belgium Bermuda Bulgaria Burkina Faso Canada China C¿te D'Ivoire Croatia Curacao Cyprus Czech Republic Egypt France Germany Ghana Gibraltar Greece Guatemala Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kenya Korea, Republic of Kosovo Kuwait Latvia Lebanon Libya Luxembourg Macao Macedonia Malaysia Maldives Mali Malta Mauritius Mexico Mongolia Myanmar Nepal Netherlands Netherlands Antilles New Zealand Nicaragua Nigeria Norway Oman Pakistan Palestine, State of Panama Peru Philippines Poland Portugal Puerto Rico Qatar Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Suriname Sweden Switzerland Taiwan Thailand Trinidad and Tobago Turkey Uganda United Arab Emirates United Kingdom Uzbekistan Venezuela Viet Nam

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    if an AAA battery is inserted in the recorder and a user attempts to start it, or if the recorder is inserted in the docking station, the recorder will display Error: 602 and fail to function for a new patient study.