Device product
SHERPA NX ACTIVE GUIDING CATHETER, 6F, 3DRC 100CM .070", REF SA63DRC. for cardiovascular use
Z-2564-2019
Product summary
- Event
- Event 82525
- Status
- Completed
- Classification
- Class I
- Quantity
- 538 units
- Official record key
device-enforcement:Z-2564-2019
Official wording
Reason: There is a potential for extensive loss of primary segment material exposing underlying stainless-steel braid wires on a subset of 6F Sherpa catheters.
Code information: GTIN: 00613994823823 All Lot/Serial Numbers
Distribution pattern: Worldwide distribution to US, Austria, Belgium, Bulgaria, Czech Republic, France, Germany, Greece, Hong Kong, Ireland, Italy, Kazakhstan, Luxembourg, Moldova, Netherlands, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Taiwan, and United Kingdom
Derived failure modes
-
Unknown
There is a potential for extensive loss of primary segment material exposing underlying stainless-steel braid wires on a subset of 6F Sherpa catheters.