iMRX MRI and Angiography System, part number 10094137 and 10094141, a surgical/interventional suite with integrated MRI and X-ray angiography. The IMRIS iMRX system is the integration of a Siemens Artis zee family (biplane or floor or ceiling mounted single plane) and the IMRIS Neuro 3T System. The System is a solution providing MRI scanning capabilities in the angiography room while retaining all functionality of Siemens Artis zee system. The System includes a traditional MRI unit that has been suspended on an overhead rail system, and is designed to operate inside RF room to facilitate intra-operative, interventional, and multi-room imaging. It also allows intra-operative or interventional imaging capabilities, while retaining all diagnostic functionality of a standard MRI system in the diagnostic room. The iMRX system allows for the sequential use of either MR imaging or X-ray fluoroscopic imaging of the head or chest for a single patient while positioned on a common fixed table either prior to, during or post clinical procedure. When these images are interpreted by a trained physician, they may assist in diagnosis during either neurovascular or cardiovascular interventionsa procedures. The iMRX system is designed such that the imaging sub-systems cannot operate simultaneously and only in accordance with their respective cleared intended use.

Z-0762-2020

November 26, 2019

Class II

Product summary

Firm
Deerfield Imaging, Inc.
Event
Event 84400
Status
Terminated
Classification
Class II
Quantity
6
Official record key
device-enforcement:Z-0762-2020

Official wording

Reason: There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible

Code information: Serial Number: 146792, 153374, 153642, 153793, 153896, 154588.

Distribution pattern: CT, MA, VA. International distribution: Australia, Canada, and Qatar.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible