Skip to content
Recall Observatory FDA recall evidence

Device product

DELFI SINGLE USE VARIFIT TOURNIQUET CUFF SINGLE PORT SINGLE BLADDER CONTOUR THIGH CUFF WITH MATHCING LIMP PROTECTION SLEEVE Rx Only, Sterile

Z-1902-2020

April 03, 2019

Class II

Product summary

Firm
Delfi Medical Innovations, Inc.
Event
Event 84741
Status
Terminated
Classification
Class II
Quantity
24 units
Official record key
device-enforcement:Z-1902-2020

Official wording

Reason: Delfi Medical Innovations Inc has discovered that tourniquets cuffs in certain lots of the VariFit Sterile Single Use Thigh Cuffs, Single Port, Single Bladder models may have been manufactured with an inadequate perimeter weld on the cuff. Once applied to a patients thigh and inflated to a pressure to occlude blood flow in the thigh distal to the cuff, after a period of time the cuff may leak to the extent that the tourniquet instrument supplying air pressure to the cuff may not be able to adequately control and regulate pressure in the cuff. This may result in some blood flow in the thigh distal to the cuff.

Code information: 18J15

Distribution pattern: US: Ohio OUS: Austrailia

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Delfi Medical Innovations Inc has discovered that tourniquets cuffs in certain lots of the VariFit Sterile Single Use Thigh Cuffs, Single Port, Single Bladder models may have been manufactured with an inadequate perimeter weld on the cuff. Once applied to a patients thigh and inflated to a pressure to occlude blood flow in the thigh distal to the cuff, after a period of time the cuff may leak to the extent that the tourniquet instrument supplying air pressure to the cuff may not be able to adequately control and regulate pressure in the cuff. This may result in some blood flow in the thigh distal to the cuff.