Device product
System AXIOM Aristos FX Plus - A dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems; and allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, and extremities. It is not designed for mammography use. Radiographic exposures may be taken with the patient in the sitting, standing, or supine positions.
Z-1765-2020
Product summary
- Event
- Event 85448
- Status
- Terminated
- Classification
- Class II
- Quantity
- 399 (wordwide)
- Official record key
device-enforcement:Z-1765-2020
Official wording
Reason: It was determined that in rare cases persons can receive low dose of unintended radiation. The system performed exposures without the exposure button being pressed. Exposure will be released for less than 500ms, therefore, the overall dose is estimated at 0.01mSv or less than 500 mGy in total.
Code information: serail number 1192
Distribution pattern: Worldwide Distribution
Derived failure modes
-
Unknown
It was determined that in rare cases persons can receive low dose of unintended radiation. The system performed exposures without the exposure button being pressed. Exposure will be released for less than 500ms, therefore, the overall dose is estimated at 0.01mSv or less than 500 mGy in total.