C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Z-0617-2020

October 21, 2019

Class II

Product summary

Firm
Integra LifeSciences Corp.
Event
Event 84216
Status
Terminated
Classification
Class II
Quantity
820
Official record key
device-enforcement:Z-0617-2020

Official wording

Reason: The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Code information: French OPS Manual P/N 60905770

Distribution pattern: US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.