Device product
ORTHOFIX Catalogue Number: ref: 99-92502, PREFIX II DIAPHYSEAL COMPLETE KIT STERILE NO SCREWS, RX ONLY, UDI: (01) 18033509855955
Z-0391-2020
Product summary
- Firm
- Orthofix Srl
- Event
- Event 84004
- Status
- Terminated
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-0391-2020
Official wording
Reason: As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Code information: Lot # B1209789 *Not marketed in US
Distribution pattern: US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel
Derived failure modes
-
Device software or design