ARCHITECT HAVAB-G Calibrator, REF 6L27-01

Z-0574-2020

November 07, 2019

Class II

Product summary

Firm
Abbott Gmbh & Co. KG
Event
Event 84238
Status
Terminated
Classification
Class II
Quantity
83 kits
Official record key
device-enforcement:Z-0574-2020

Official wording

Reason: Abbott has identified that Calibrator lot 08219BE00 is potentially impacted by an increase in Relative Light Units (RLUs) up to 94%. It has been confirmed that the increase in the calibrator RLU values may cause a decrease in the control values and patient results. Patient values generated using the impacted calibrator lot 08219BE00 may have changed interpretation from reactive to falsely non reactive. Results impacted may present as a (falsely) non reactive result that fall in the range between 0.50 to 0.99 S/CO.

Code information: lot number 08219BE00

Distribution pattern: US in the states of: AZ, CA, CO, CT, FL, KY, LA, MA, MD, MI, MN, MO, NJ, NY, OH, OR, RI, SC, TN, TX, WA, WV

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Abbott has identified that Calibrator lot 08219BE00 is potentially impacted by an increase in Relative Light Units (RLUs) up to 94%. It has been confirmed that the increase in the calibrator RLU values may cause a decrease in the control values and patient results. Patient values generated using the impacted calibrator lot 08219BE00 may have changed interpretation from reactive to falsely non reactive. Results impacted may present as a (falsely) non reactive result that fall in the range between 0.50 to 0.99 S/CO.