Skip to content
Recall Observatory FDA recall evidence

Device product

SIGNA Premier Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.

Z-0874-2020

December 23, 2019

Class II

Product summary

Firm
GE Healthcare, LLC
Event
Event 84655
Status
Terminated
Classification
Class II
Quantity
62 (30 US, 32 OUS) total devices
Official record key
device-enforcement:Z-0874-2020

Official wording

Reason: It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.

Code information: Mfg. Lot or Serial# System ID 33-4-1031 415723SHMR19 To be provided RTD01204 To be provided DPN88672

Distribution pattern: Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0