Device product
Genius 2 Tympanic Thermometer OEM Tympanic Item Code: 303063
Z-0993-2020
Product summary
- Event
- Event 84202
- Status
- Terminated
- Classification
- Class II
- Quantity
- in total 37567 units US; 335573 OUS
- Official record key
device-enforcement:Z-0993-2020
Official wording
Reason: The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.
Code information: All product manufactured after December 10, 2017; Serial Numbers greater than or equal to N17606406
Distribution pattern: Worldwide Distribution - US Nationwide and Foreign in the countries of: Australia Austria Belgium Canada Canary Islands Chile Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana French Polynesia Gabon Germany Gibraltar Greece Guadeloupe Hong Kong Iran, Islamic Republic Of Ireland Israel Italy Kuwait Liechtenstein Luxembourg Martinique Mayotte Mexico Netherlands New Caledonia New Zealand Norway Panama Philippines Poland Portugal Puerto Rico Qatar Reunion Romania San Marino Saudi Arabia Singapore Slovakia Slovenia South Africa South Korea Spain Swaziland Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom
Derived failure modes
-
Unknown
The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.