Skip to content
Recall Observatory FDA recall evidence

Device product

Norm-O-Temp model 111W

Z-2356-2019

June 20, 2019

Class II

Product summary

Firm
Cincinnati Sub-Zero Products LLC, a Gentherm Company
Event
Event 83376
Status
Terminated
Classification
Class II
Quantity
1063
Official record key
device-enforcement:Z-2356-2019

Official wording

Reason: The manual does not provide instruction stating that exceeding 40 Celsius for extended periods can cause tissue damage.

Code information: Serial number Range: 091-N6669-192-N7901 UDI: 10613031861563, 10613031861570, 10613031861587, 10613031861617 Part # 86156, 86157,86158, 86161

Distribution pattern: Ship-To State AL AR AZ CA CO CT DC DE FL GA HI IL IN KS KY LA MA MD ME MI MN MO NC ND NE NH NJ NM NY OH OK PA RI SC TN TW TX UT VA WA WI WV WY Ship-To Country United States Taiwan Egypt United Arab Emirates Israel Pakistan Turkey Colombia Thailand Kuwait United Kingdom Mexico Netherlands Canada Ecuador South Korea Germany Georgia China Panama Japan Malaysia Hong Kong Austria Chile Mongolia Morocco Australia Indonesia Venezuela Singapore

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0