AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic, Type number: 6010/200 - Product Usage: DR 800: Agfa s DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: positioning fluoroscopy procedures, gastro-intestinal examinations, urogenital tract examinations, and angiography. It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts. The DR 800 is not intended for mammography applications

Z-2202-2020

April 06, 2020

Class II

Product summary

Firm
Agfa N.V.
Event
Event 85599
Status
Terminated
Classification
Class II
Quantity
13 units
Official record key
device-enforcement:Z-2202-2020

Official wording

Reason: Under specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possible

Code information: Serial Numbers: 0925, 0945, 0948, 0949, 1003, 1022, 1116, 1136, 1161, 1229, 1239, 1401, and 1404

Distribution pattern: US Nationwide distribution including in the sates of FL, IL, IN, LA, MA, MO, OH, SC, VA, WI.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Under specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possible