Skip to content
Recall Observatory FDA recall evidence

Device product

Dexcom G6 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.

Z-1562-2020

February 03, 2020

Class II

Product summary

Firm
Dexcom Inc
Event
Event 84891
Status
Terminated
Classification
Class II
Quantity
275583
Official record key
device-enforcement:Z-1562-2020

Official wording

Reason: The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.

Code information: Device Listing Number: D319131 Model: STS-OR-003 Description: G6 Sensor Kit 3 Pack GUDID Registered GTIN: 50386270000250 Lot Number: All Model: STS-OE-003 Description: G6 Sensor Kit 3 Pack Retail GUDID Registered GTIN: 00386270000866 Lot Number: All Model: STS-OM-003 Description: G6 Sensor Kit 3 Pack Medicare GUDID Registered GTIN: 00386270001047 Lot Number: All Model: STS-OM-003 Description: G6 Sensor Kit 3 Pack Pro Q GUDID Registered GTIN: 70386270000254 Lot Number: All

Distribution pattern: World-wide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0