Device product
BardPort M.R.I. Implantable Port with Attachale 9.6 F Open-Ended Single-Lumen Venous Catheter, REF: 0602680, UDI:(01)00801741025679 Product Usage: Bard implantable ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.
Z-1576-2020
Product summary
- Event
- Event 84914
- Status
- Terminated
- Classification
- Class II
- Quantity
- 2493 Catheters
- Official record key
device-enforcement:Z-1576-2020
Official wording
Reason: Firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.
Code information: Lot # RECU2404, RECU1067, RECR1439, RECT0078
Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS,KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT,VA, VT, WA, WI, WV, WY and countries of Brazil, Canada, EMEA, Korea, Mexico, New Zealand.
Derived failure modes
-
Unknown
Firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.